EES-101 Series Ultrasonic Flowmeter
Revision: 3.0.1
Page 26 of 65
Empty Pipe Setup
This parameter is used to overcome the possible
problems that usually show up when the pipe being
measured is empty. Since signals can be transmitted
through the pipe wall, the flow meter may still read a
flow while measuring an empty pipe. To prevent this
condition from happening, you can specify a value.
When the signal quality falls below this value, the
measurement stops automatically. If the flow meter is
already able to stop measuring when the pipe is empty,
a value in the range of 30 to 40 should also be entered
in this window to ensure no measurement when the
pipe is empty.
Measurement Units Options
Select the measurement unit as follows:
0.
Metric
1.
English
Flow Rate Units Options
The following flow rate units are available:
0.
m3
Cubic Meters
1.
1
Liters
2.
gal
USA Gallons
3.
ig
Imperial Gallons
4.
mg
Million Gallons
5.
cf
Cubic Feet
6.
bal
USA Barrels
7.
ib
Imperial Barrels
8.
ob
Oil Barrels
The following time units are available:
/ Day
/ Hour
/ Min
/ Sec
Factory default is Cubic Meters/hour
Totalizer Units Options
Select totalizer units. The available unit options are as
same as those found in Window M31. The user can
select units as their requirement. Factory default is
Cubic Meters.
Menu
2
9
Empty Pipe Setup [29
0
Menu
3
0
Measurement Units In
0. Metric
Menu
3
1
Flow Rate Units [31
m3/h
Menu
3
2
Totalizer Units [32
0. Cubic Meter (m3)