EM TEST
PFS 200 Series
Manual for Operation
V 3.08
18 / 34
3.9.
Setup
This menu helps the user to define the configuration of the PFS 200.
SETUP
F1 : Change language /
Sprache ändern
F2 : LCD backlighting
F3 : Interfaces
F4 : Keyboard-Beeper
F5 : Timer
F6 : Set voltage
F1
F2
F3
F4
F5
F6
F7
F1 Change language
The user can chose between two languages, German and English.
F2 LCD backlighting
With F2 the backlighting can be switched On or Off. Additionally the Auto Off function can be programmed to
switch off the backlighting after a specified time the generator has not been in operation (1- 30min).
F3 Interfaces
This menu will help the user to define the status of the integrated serial and parallel interfaces, e.g. the baud rate
of the Serial Interface or the address of the IEEE interface.
F4 Keyboard-Beeper
F1 is the selector for the beeper ON/OFF mode.
The beeper is always on when a test routine is finished. To indicate that a running test is finished the beeper
sounds 3 times.
F5 Power ON counter
Pressing of F5 will show the total operating time of the test equipment.
F6 Set voltage
For control of an external power supply source an analogue control voltage can be programmed (0-10V dc).
The operator can specify the following parameters:
F1:
Max output voltage
of the variac. The analogue reference value for this is 10V.
F2:
Rated output voltage
for the device under test. This voltage shall be specified by the operator and
depends on the type of equipment under test. The variac normally is automatically set to this output voltage.