EM TEST
PFS 200 Series
Manual for Operation
V 3.08
14 / 34
3.6.
Ford test routines as per ES-XW7T
The setting of Ford parameters on device is only
possible if in menue
F7 Service / F3 Setup / F6 Set voltage
the parameter
F2 max. DC source voltage
is connected to a source with
at least
13.5V voltage
.
3.6.1. Power Dropout ( CI 260 )
This test represents intermittent power interruptions and switch bounce.
Page 4
( menus)
Ford
F1 : Ford Dropout High
F2 : Ford Dropout Low
F3 : Ford Dropout Single
F4 : Ford Dip
F1
F2
F3
F4
F5
F6
F7
3.6.2. Ford Voltage Drop High CI 260-A
Parameters:
t : 100
s , 200
s , 400
s , 700
s ,
1ms, 3ms , 7ms, 10ms, 12ms, 18ms,
20ms, 50ms
t1 : 0.1s to 99.0s
tri Auto / Manual
n :
Rise and fall-times are approximately 10
s
3.6.3. Voltage Drop Low CI 260-B
Parameters:
ti : 0.1s to 99.0s
t : 100
s , 200
s , 400
s , 700
s , 1ms,
3ms, 7ms, 10ms, 12ms, 18ms, 20ms,
50ms
t1 : 0.10s to 99.00s
tri : Auto / Manual
n :
Rise and fall-times are approximately 10
s