10.1.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
Quick Start Menu
While a test is running the user can select parameters. The selected parameter then can be changed online with
the inc/dec knob.
10.1.2.
Standard test Routine
The user can select preprogrammed standard test routines.
The Surge Standard menu offers a list of standards.
The name of the standard defines the characteristic of
the rulers that is given in the application of the
standard.
The standard name has the character:
[Standard Family], [Standard number], [year]), [Application],
where
Standard Family
IEC:
International Electrotechnical Commission
EN:
European Committee for Standardization (CEN)
Standard Number
Indicates the official standard number defined in the standard.
Year
Year of standard publication. The year indicates the different standard publication. The year is used for follow the
definitions specified in this publication.
Application
Specifies different applications or EUT characteristics
-
One level: only one test level will be in this test. The user can select the test level (level 1 to 4)
-
All levels: The test procedures start at the lowest level and increase to the level specified.