EM TEST
Compact NX5
Operating Manual
V 1.06
85 / 143
9.2.1.2. Signal Trigger
The Signal Trigger occurs by pressing the blinking
START / PAUSE
button or with an external trigger
positive slope signal at the rear side BNC
Ext. Trigger
plug.
The start trigger offers the following status:
-
Automatic
- At
Start
(starts after the trigger)
If the start trigger is not set to Automatic, the trigger indicates with “
Waiting for trigger
”
the recommended
trigger signal.
The waiting time for a trigger signal is endless if no trigger timeout is enabled.
External trigger timeout
This function stops the test after the defined timeout time
and indicates the message,
“
Error trigger unit timeout
”.
For enable the function External trigger timeout select from
the Home screen:
Menu / Setup / General / Process & sync
Start Trigger flowchart
9.2.1.3. EUT Monitor
The EUT monitor 1 and EUT Monitor 2 input is used for control the compaxt NX device according the EUT
behavior. Each input is programmable to have different events as:
Disabled
No function
Notify
Makes a mark in the NX software.
The report will make a time stamp and notes the actual generator settings.
Stop
Stops the test immediately. It is not possible to continue the test.
Pause
(EUT monitor 1)
Break the test and go to pause status. For continue the user must press the
START/PAUSE button or continue in iec.control software.
NOTE: The EUT monitor input 1 must go to
open status before press continue
Hold
(EUT monitor 2)
Keep in pause status as long the EUT monitor 2 input is grounding.
After release the break, the test will continue automatically.
Содержание compact NX5 bp-1-300-16
Страница 142: ......
Страница 143: ...EM TEST Compact NX5 Operating Manual V 1 06 143 143 16 4 Main diagram high voltage connection ...