Power frequency Magnetic Field as per IEC 61000-4-8
Main Menu Magnetic Field separated to
“Power frequency” and “Pulsed”
Required Device settings to perform a power frequency magnetic field test
For perform a 50/60 Hz magnetic field test, the NX5 generator must be configured with the used hardware in the
menu:
Menu / SETUP / Equipment
1.
Variable transformer
:
- variac NX1-260-16
- variac NX1-260-32
2.
Magnetic current transformer
- MC 26100
H-field 75 to 1100 A/m
- MC 2630 (1 A range)
H-field
1 to 9 A/m
- MC 2630 (30 A range) H-Field
9 to 75 A/m
Note: The MC 2630 has two ranges and therefore the NX generator list this current-transformer in the
configuration for (1 A) and (30 A) range. Set the range switch of the MC 2630 to the correct position before
start the test.
Menu / SETUP / EXTERNAL COUPLERS
3.
Magnetic field antenna
:
- MS 100N
Set the used equipment to default status