EM TEST
Compact NX5
Operating Manual
V 1.06
51 / 143
Path
Vnom
(controlled at PF2)
PF2
Any voltage (example 230 V- 10%)
The variac voltage is set to the following
voltages:
Power Fail:
settled DIP voltage
Power magnetic Field:
0 V
All other
phenomenon’s: Vnom (PF2)
NOTE: PF2 is used for operate a EUT with different voltage than
the actual mains voltage.
For Power Fail tests only PF2 mode with voltage interrupt
tests are possible.
4.3.5.3.
EUT Setup / LIMIT
LIMIT
Defines the limit for the voltage and current rms measurement for stop the test.
-
Enable the box
for activate and set the limit
- Current limit will trip if the current is above the settled limit and will
break / stop the test
- Voltage limit will trip if the rms voltage is above the settled limit and will
break / stop the test
Enabled
(white)
Disabled
(grey)
Current
: Range 0 to 50 A
no function
Voltage
: Range 0 to 300 V
no function
Internal hardware current limiter
The built in internal hardware limiter operates as follow. In case of an overcurrent the
TEST ON
is switched off,
and the user has to confirm a message.
16 A models
I < 23 A
No action
23 A
≤ I < 40 A
Fail after 45 s
40 A
≤ I < 50 A
Fail after 30 s
50 A
≤ I
Fail after 5 s
32 A models
I < 46 A
No action
46 A
≤ I < 80 A
Fail after 45 s
80 A
≤ I < 100 A
Fail after 30 s
100 A
≤ I
Fail after 5 s