4.3.5.1.
EUT Setup / TEST
TEST
Power configuration for the actual connected EUT
Supply
:
AC
DC
Coupling
:
Power lines
Data lines
Phase Filter
:
Select the used lines
- 1-phase
(L, N, PE)
- 3-phases
(L1, L2, L3, N, PE)
Coupling
Select the used coupling Mode to the selected coupling device
- 1 phase for NX5 device
- 1 or 3 Phase for coupling NX
Supply
AC
Coupling
Power Lines
1 Phase
L, N, PE
with PE
L, N
no PE
3 Phases
L1, L2, L3, N, PE
Star with PE
L1, L2, L3, N
Star no PE (PEN)
L1, L2, L3, PE
Delta with PE
L1, L2, L3
Delta np PE
L1, L2, PE
Interphase
Lx, Lx, …
other combination
Supply
DC
Coupling
Power Lines
Plus, Minus, PE
with PE
Plus, Minus
no PE
4.3.5.2.
EUT Setup / SUPPLY
SUPPLY
Defines the default supply path to the EUT and the nominal voltage. For DC application the user
needs two DC sources for PF1 and PF2 voltage each.
Path
:
PF1
Mains voltage from the power grid
PF2
Voltage from the variac or tapped transformer
Vnom
:
Mains voltage or defined voltage
Path
Vnom
(mains from the grid)
PF1
100 V, 115 V, 230 V, or any mains voltage
NOTE: PF1 is the default setting for general application for Burst,
Surge and Power Fail testing.
The compact NX generator uses this V
nom
voltage as
100% value for Dips
test as per IEC 61000-4-11.