EN
Problem/Possible Cause
*Auto-
Reset
Function
Action
Error Codes: 302
hex
(770
dec
); 303
hex
(771
dec
)
—sensor EEPROM read error
Sensor EEPROM with sensor calibration
values could not be read correctly after
the sensor was connected or after the
BioProTT
TM
FlowMCP was started.
1.
Check if the sensor is connected properly to the
BioProTT
TM
FlowMCP and ensure that the sensor is not
damaged.
2.
Check if there are not electromagnetic disturbances on the
sensor cables (e.g. disturbances from a pump).
3.
Disconnect the sensor.
4.
Reconnect the sensor again.
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
Error Codes: 401
hex
(1025
dec
)
—signal processing FPGA test cycle failed
During an internal signal processing
test of the FPGA component on the
flow measurement board, a failure was
detected.
Electromagnetic disturbances could
lead to this failure situation.
If auto-reset function is not set:
1.
Check if there are any electromagnetic disturbances that
might influence the system. Also check the line grounding.
2.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
3.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
If the auto-reset function is set:
The BioProTT
TM
FlowMCP resets the board to clear the error.
→
If the error is no longer present, then the device can be used
like normal.
→
If the problem persists, return the device for servicing.
Error Codes: 402
hex
(1026
dec
)
—sensor wire break of measurement signal lines detected
The sensor cable (including the sensor
extension cable) is damaged (e.g. wire
break) or the tube is not completely
filled with liquid.
1.
Check if the sensor is connected properly to the
BioProTT
TM
FlowMCP and ensure that the sensor cable and/or
the sensor extension cable are not damaged.
2.
Check if the tube is completely filled with liquid.
3.
Power off the device (disconnect the power supply).
4.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
BioProTT™ FlowMCP
Page 58
Copyright em-tec GmbH
D214-703 BioProTT™FlowMCP – User Manual – V4.0