EN
Problem/Possible Cause
*Auto-
Reset
Function
Action
Error Codes: 202
hex
(514
dec
)
—
run-time self-test on flow measurement board failed during memory check
Internal failure (ROM failure) during
run-time.
If auto-reset function is not set:
1.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
2.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
If the auto-reset function is set:
The BioProTT
TM
FlowMCP resets the board to clear the error.
→
If the error is no longer present, then the device can be used
like normal.
→
If the problem persists, return the device for servicing.
Error Codes: 203
hex
(515
dec
); 204
hex
(516
dec
); 205
hex
(517
dec
); 206
hex
(518
dec
); 207
hex
(519
dec
); 20E
hex
(526
dec
)
—
run-time self-test on flow measurement board failed during flow measurement board voltage checks
Internal voltage failure was detected
on the flow measurement board during
run-time mode.
Ensure the power supply for the BioProT
TTM
FlowMCP is within
the specified operating range.
1.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
2.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
Error Codes: 208
hex
(520
dec
)
—
run-time self-test on flow measurement board failed during flow measurement board temperature checks
The temperature inside the device is
too high.
Ensure the ambient temperature is within the specified opera
-
ting range. If the temperature is too high, cool the device down.
1.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
2.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
BioProTT™ FlowMCP
Page 56
Copyright em-tec GmbH
D214-703 BioProTT™FlowMCP – User Manual – V4.0