Problem/Possible Cause
*Auto-
Reset
Function
Action
Error Codes: 1
05
hex
(261
dec
); 106
hex
(262
dec
)
—
start-up temperature check failed
The temperature inside the device is
too high.
Ensure the ambient temperature is within the specified opera
-
ting range. If the temperature is too high, cool the device down.
If auto-reset function is not set:
1.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
2.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
If the auto-reset function is set:
The BioProTT
TM
FlowMCP resets the board to clear the error.
→
If the error is no longer present, then the device can be used
like normal.
→
If the problem persists, return the device for servicing.
Error Codes: 201
hex
(513
dec
);
202
hex
(514
dec
)
—
run-time self-test on flow measurement board failed during memory check
Internal failure (ROM failure) during
run-time.
If auto-reset function is not set:
1.
Power off the device (disconnect the power supply) or
perform a reset of the flow channel.
2.
Power on the device again (reconnect the power supply).
→
If the error is no longer present, the device can be used like
normal.
→
If the problem persists, return the device for servicing.
If the auto-reset function is set:
The BioProTT
TM
FlowMCP resets the board to clear the error.
→
If the error is no longer present, then the device can be used
like normal.
→
If the problem persists, return the device for servicing.