Maintenance
69
10.1.10
Cleaning the connecting hose between the furnace and metal filter
Clean the hose connection from the dust filter to the furnace on the analyzer as follows:
1. Open the machine door (see Pos.
11
in
”).
The following figure shows the hose connection from the dust filter to the furnace of the
analyzer:
Fig. 28: Hose connection from the dust filter to the furnace
2. Unscrew the hose nuts on the filter connector and on the bottom furnace locking device (
1
).
3. Pull the hose (
2
) off.
4. In der ELEMENTS software, start the vacuum (
“Analyzer status” window under “User
service”) or use an external vacuum cleaner to remove falling particles while cleaning.
5. Clean the hose with a pipe cleaner (Part No. 70002) or with a micro-brush (Part No. 88400-
0501).
6. Replace the hose (Part No. 51100-2040) if it can no longer be cleaned or leaks.
7. Insert the hose (
2
) and secure the hose nuts (
1
).
8. Close the machine door.
The hose connection from the dust filter to the furnace has been cleaned.
1
2