Installation
25
4.1 Setting up
CAUTION
C15.0000
Risk of posture problems
Lifting the HTFr machine unaided
−
The weight of the HTFr machine can cause injuries resulting from
posture problems due to an unhealthy posture or inattentiveness when
lifting.
•
Only lift the machine with the help of a second person.
•
Ensure you maintain a healthy posture when lifting the machine.
CAUTION
C16.0092
Machine falling down
Incorrect installation or insufficient working space
– Due to its weight, the machine can cause injuries if it falls down.
•
Only operate the machine in a sufficiently large, robust, non-slip and
stable working area.
•
Ensure that all machine bases are placed securely on the surface.
The space required to set up the ELEMENTRAC CS-d with balance and PC is approx. 190 cm
× 75 cm (width x depth), without balance and PC approx. 100 cm × 75 cm.
The machine bench must be able to safely withstand the weight of the
machine (see “
”). The power plugs must be easily accessible at all times. The surface of the machine
bench must be non-slip and fire-resistant.
Set the machine up as follows:
1. Set the machine up on the prepared workstation.
2. Make sure that the HTFr is to the right of the analyzer when viewed from the front.
3. Ensure that the power plugs on the analyzer and of the HTFr are freely accessible at all
times.
4. Make sure to set up both machines in a way that does not block the fans.
5. Set up the PC, monitor, keyboard and balance in the desired place.
6. Ensure that the machine is secure and stable.
7. Ensure that the surface of the workstation is non-slip and fireproof.
8. Make sure that no crucibles are able to roll down from the surface.
9. Ensure that there are no flammable objects or materials (work papers, files) near the
analyzer.