Maintenance
45
10.1.3.1
Filling the reagent tube on the SO
3
filter
Proceed as follows to replace the filling in the reagent
tubes on the SO
3
filter (see Pos. (
3
) in
“):
1. Turn the main switch on the analyzer to Position 1.
2. Remove the reagent tube, see
“.
3. Remove the used filling from the reagent tube.
4. Dispose of the reagent tube filling in accordance with
local regulations and provisions.
5. Ensure that the reagent tube is undamaged.
If it is damaged or too contaminated, use a new
reagent tube (Part No. 88400-0422).
6. Fill the reagent tube with 3 to 4 cm of glass wool (Part
No. 90331).
7. Fill the reagent tube with 10cm of cellulose (Part No.
90341).
8. Insert the filled reagent tube, see
.
The filling of the reagent tube on the SO
3
filter has
been replaced.
10cm
90341
3-4cm
90331