Maintenance
42
CAUTION
C25.0090
Risk of eye injury
Chemicals
−
When changing the chemicals, minute particles of chemicals may be
suspended in the air and cause burns to eyes.
•
Always wear protective goggles when working with chemicals.
•
Comply with the safety data sheets for the chemicals used.
The reagent tubes of the analyzer must not be removed during operation.
The following diagram shows the filters on the front of the analyzer with the machine door open:
Fig. 8: Front of the analyzer with the door open
Position
Component
1
Moisture filter
2
Inlet gas purification
3
SO
3
filter
Proceed as follows to remove a reagent tube on the front of the analyzer:
1. Turn the main switch on the analyzer to Position 1.
2.
On the “Status analyzer” page of the ELEMENTS software, press the “Change reagents”
button to release the pressure.
3
2
1