Servicing
23
Proceed as follows to check the chain tension:
1.
Press a number of crucible trays in the back row down (
1
) to hold them in
their current position (see Fig. 11: Crucible chain).
2.
Move the front crucible trays (
2
) to the right and left (see Fig. 11: Crucible
chain). The crucible trays should have approx. 1 mm of play.
NOTICE
Blockage
Incorrect tension
Excessive tension can lead to a blockage when the crucibles are moved
by the motor. If the chain has too much play, bent elements on the chain
can block the cogwheel support. A chain play of approx. 1 mm should be
maintained (see description above).
Proceed as follows to adjust the chain tension:
1.
Release the screw (
1
) on the underneath of the sample changer platform
(see Fig. 12: Adjusting the chain tension).
2.
Move the screw (
1
) to and fro to adjust the chain tension (see Fig. 12:
3.
Check the chain tension again.
The chain tension has now been adjusted.