BIOTOPUS II
- Operator manual
Page 23
probe by adjusting the proper screw (in
correspondence to the plug) on the top
of the interface (black box):
use left screw to adjust the probe con-
nected to the left plug, right screw to
adjust the probe connected to the right
plug.
Low beeping frequency shows a value
far away from Optimal Point, high bee-
ping frequency shows a value closer to
the Optimal Point. On the display the va-
lues should come to the Optimal range
(440 - 460).
Once the Optimal range is achieved,
beeping will stop.
Press CONTINUE to fi nish the calibration
process.
Содержание BIOTOPUS II
Страница 2: ...BIOTOPUS II Operator manual Page 2...
Страница 5: ...BIOTOPUS II Operator manual Page 5 Figure 1 1...
Страница 6: ...BIOTOPUS II Operator manual Page 6...
Страница 30: ...BIOTOPUS II Operator manual Page 30...
Страница 31: ...BIOTOPUS II Operator manual Page 31 ADAVANCED OPERATION...
Страница 46: ...BIOTOPUS II Operator manual Page 46 MISCELLANEOUS...
Страница 68: ...BIOTOPUS II Operator manual Page 68...
Страница 72: ...BIOTOPUS II Operator manual Page 72...