9
2. When the display shows measurement result that you feel are satisfactory for
use as your target values, press
H
key to hold these results.
3. Press
key to enter the color difference measurement mode. If you have
previously set a target, the display shows the difference between the new
measurement and this previously set target.
4. Press
SET
key to enter the
SET
mode, the annunciator
“
SET
REF
” and the
measurement result that you held at step 2 above now blink on the display.
If you wish to cancel without store the blinking values as your new target,
press
SET
key again.
5. Press
key to store the measured value into the target memory, so that they
become the current target values. The displayed values accordingly change
“
+0.0
” (Where the number of digits on each line varies according the display
mode), since the measured values and target values are now identical.
6. Press
H
key again to exit
Hold
mode and back to the color difference
measurement mode. In this mode, you can review the target values at any
time by holding down the
key. The values remain on the screen until you
release the key.
5-3-2 Manually keying in the target values
Note that the dominant wavelength, the excitation purity and the
colorimetric purity cannot be used to set the target value.
1. Press
MODE
key to select the desired (Ev, x, y), (Ev, u
’, v’) or (X, Y, Z)
display mode.
2. Press key to enter the color difference measurements mode.