10.2 Cable connection, grounding
For interconnecting with another devices there it is necessary to use only
shielded cables. On device side shielding of the cable must be attached with metal
body of cable gland (according to Fig. 30). All cable glands of the device are
connected reciprocally. It means that shielding of all cables coming into the device is
interconnected. Thereby is ensured high resistance against electromagnetic
disturbance.
Temperature sensor and external pressure transducer (if included) is also
equipped with cable whose shielding is attached to metal body of cable gland. Metal
part of temperature sensor is insulated. Metal body of pressure transducer is
connected with cable shielding.
During installation of the device and connecting of shielding is important to
avoid creating of ground loop.
Device is not necessary to ground.
For cable connection (size of conductor 0,5 – 1,5 mm
2
) there are used clamps
mounted in the device, in whose nearness there is printed on the board description of
signal which is assigned to each clamp (see Fig. 12). Before connecting cables it is
neccessary first to apply wire-end ferrule on the stripped end of the wire and force on
them by pliers which are supplied by the manufacturer of wire-end ferrules. Wires
ended by ferrules it is possible to plug in to the clamps without need of special tool.
During wire dissconnecting it is neccessary to slightly push nose of the clamp and
carefuly take out the wire.
cable type cable diameter Recommended cable type
Pulse input
Shielded 2
wire cable
4 – 6,5 mm
Unitronic LiYCY 2 x 0.25 Lappkabel Stuttgart
SRO 2.22
Č
SN347761 Kablo Velké Mezi
ř
í
č
í
External power
feeding
Shielded 2
wire cable
4 – 6,5 mm
Unitronic LiYCY 2 x 0.75 Lappkabel Stuttgart
Pulse outputs
Shielded 6
wire cable
4 – 6,5 mm
Unitronic LiYCY 6 x 0.25 Lappkabel Stuttgart
SRO 6.22
Č
SN347761 Kablo Velké Mezi
ř
í
č
í
RS-232
connection
Shielded 4
wire cable
4 – 6,5 mm
Unitronic LiYCY 4 x 0.25 Lappkabel Stuttgart
RS-485
connection
Shielded 4
wire cable
4 – 6,5 mm
Unitronic LiYCY 4 x 0.34 Lappkabel Stuttgart
Tab. 11 Recommended cable types
Temperature resistance of SRO cable type is given by manufacturer in range -
5°C - +45°C. Manufacturer Lappkabel Stuttgart is guarantee resistance on their
product Unitronik LiYCY for moving lead-wires in range -5°C - +70°C and for
immovable lead-wires in range -30°C - +80°C. From these values ensue also
temperature conditions for mounting of the device.
Содержание miniELCOR
Страница 13: ...miniELCOR 6 Fig 1 Volume and energy calculations Scheme...
Страница 19: ...miniELCOR 12 Fig 5 Security marks device without SCR encoder Fig 6 Security marks of miniELCOR SCR...
Страница 50: ...miniELCOR 43...
Страница 55: ...miniELCOR 48 Fig 22 Basic navigation from the initial screen FW 2 xx...
Страница 65: ...miniELCOR 58 Fig 26 Mounting of miniElcor to the mounting plate Fig 27 Mounting on the pipeline...
Страница 69: ...miniELCOR 62 Fig 30 Shielding connection in bushing 12 35...
Страница 84: ...miniELCOR 77 Fig 33 Base device parameters Fig 34 Data archive...
Страница 100: ...miniELCOR 93 Fig 50 Wizard for mathematical expression Fig 51 Equation for primary volume pulses and output connection...
Страница 111: ...miniELCOR 104 Fig 64 Switch of communication protocol in the device...
Страница 114: ...miniELCOR 107 Fig 66 Fig 67...
Страница 119: ...miniELCOR 112 Fig 70 Fig 71...
Страница 120: ...miniELCOR 113 Fig 72 Fig 73...
Страница 121: ...miniELCOR 114 Fig 74 Fig 75...
Страница 122: ...miniELCOR 115 Fig 76...
Страница 125: ...miniELCOR 118...