miniELCOR
4
1.2.3 Volume correction at measurement conditions
Device enables to compensate gasmeter error according to predefined
correction curve from gasmeter test certificate. This function and parameters V
c
can
be activated only by manufacturer or by Acreditive service to ensure that used
gasmeter correction curve in dependance on flowrate Q
m
is valid within working
conditions.
Error of measurement is corrected by usage of function f(Q
m
). For corrected
volume is:
V
c
= V
m
x f(Q
m
)
where
V
c
... Corrected volume at measurement conditions
V
m
... Primary
volume
Q
m
... Primary
flowrate
Linear interpolation method is used for getting values between calibration
points. File with correction values is to be inserted into device with help of service
programme 22. Information about insertion of correction curve into device is logged
in setup archive.
The principle of volume calculation are seen on Fig. 1
Condition for usage of volume correction.
1. Correction is used only in case that gasmeter transmits at least 10 pulses
per second resulting in usage only HF sensors.
2. Under Q
min
correction is not applied and over Q
max
value of correction
coefficient given for Q
max
will be used.
Conversion of volume on energy (*)
Device enables to calculate consumpted quantity of gas directly in energy form.
This conversion uses value of combustion heat H
s
. Calculation is made with
adding of differences dV
b
( and dV
bs)
multiplied by actual value of combustion heat
H
s
.
dE=H
s
x dV
b
, dE
s
=H
s
x dV
bs
Two other counters ( energy counter E and estimated energy counter E
s
) are
dedicated for measurement in configurable energy units: MJ, kWh, Btu.
Note :
No conversion of absolute counter value (E or Es) is accomplished after
change of units. Following increases are added already respecting new units.
Principle diagram of energy calculation is drawn at Fig. 1
Содержание miniELCOR
Страница 13: ...miniELCOR 6 Fig 1 Volume and energy calculations Scheme...
Страница 19: ...miniELCOR 12 Fig 5 Security marks device without SCR encoder Fig 6 Security marks of miniELCOR SCR...
Страница 50: ...miniELCOR 43...
Страница 55: ...miniELCOR 48 Fig 22 Basic navigation from the initial screen FW 2 xx...
Страница 65: ...miniELCOR 58 Fig 26 Mounting of miniElcor to the mounting plate Fig 27 Mounting on the pipeline...
Страница 69: ...miniELCOR 62 Fig 30 Shielding connection in bushing 12 35...
Страница 84: ...miniELCOR 77 Fig 33 Base device parameters Fig 34 Data archive...
Страница 100: ...miniELCOR 93 Fig 50 Wizard for mathematical expression Fig 51 Equation for primary volume pulses and output connection...
Страница 111: ...miniELCOR 104 Fig 64 Switch of communication protocol in the device...
Страница 114: ...miniELCOR 107 Fig 66 Fig 67...
Страница 119: ...miniELCOR 112 Fig 70 Fig 71...
Страница 120: ...miniELCOR 113 Fig 72 Fig 73...
Страница 121: ...miniELCOR 114 Fig 74 Fig 75...
Страница 122: ...miniELCOR 115 Fig 76...
Страница 125: ...miniELCOR 118...