18 Final verification of the device after replacement
of sensor/transducer or adding of digital
transducer
As a final step is recommended to proceed verification of the device by the
help of selfdiagnostic function. Click on „
MENU- Setup/Status (diagnostics) of
device/from Device
“. New window will be opened which describes actual status of
device (see Fig. 75). In the column „
State of last test
“ must not be any warning or
error! If device announce warning or error, click on „
Devicee test
“ button. In case
that the problem continues, contact manufacturer’s technical support.
If everything is all right and still warning or error message is displayed in
„
Summary status
“ column, click on „
Clear summ. Status
“ thereby all historical
warnings and errors occured before replacement or adding of transducer will be
cleared.
Fig. 69
Содержание miniELCOR
Страница 13: ...miniELCOR 6 Fig 1 Volume and energy calculations Scheme...
Страница 19: ...miniELCOR 12 Fig 5 Security marks device without SCR encoder Fig 6 Security marks of miniELCOR SCR...
Страница 50: ...miniELCOR 43...
Страница 55: ...miniELCOR 48 Fig 22 Basic navigation from the initial screen FW 2 xx...
Страница 65: ...miniELCOR 58 Fig 26 Mounting of miniElcor to the mounting plate Fig 27 Mounting on the pipeline...
Страница 69: ...miniELCOR 62 Fig 30 Shielding connection in bushing 12 35...
Страница 84: ...miniELCOR 77 Fig 33 Base device parameters Fig 34 Data archive...
Страница 100: ...miniELCOR 93 Fig 50 Wizard for mathematical expression Fig 51 Equation for primary volume pulses and output connection...
Страница 111: ...miniELCOR 104 Fig 64 Switch of communication protocol in the device...
Страница 114: ...miniELCOR 107 Fig 66 Fig 67...
Страница 119: ...miniELCOR 112 Fig 70 Fig 71...
Страница 120: ...miniELCOR 113 Fig 72 Fig 73...
Страница 121: ...miniELCOR 114 Fig 74 Fig 75...
Страница 122: ...miniELCOR 115 Fig 76...
Страница 125: ...miniELCOR 118...