background image

SERVICE MANUAL

 

 

 

 

- 24 - 

 

 

 

 

SM-eVENT-GB-3_8-2006.doc

 

 

 

15. LIST OF SPARE PARTS  

 

1.  L screwing 8/6-1/4   

 

025500122 

2.  Reduction 3/8F-1/4F 

 

025500123 

3.  Plug 1/8 M   

 

 

025400029 

4.  Check valve 3/8M-3/8M 

 

025300007 

5.  Pressure tank complete 3BA-295 

603011295 

6.  Hose PUR 8/6 

 

 

 

7.  Filter (D) 5 µm 

 

 

025200071 

8.  Regulator complete 4BA-338 

604011338 

9.  Filter (C) 40 µm 

 

 

025200070 

10.  Filter complete 4BA-307 

 

604011307 

11.  Hose PUR 10/7 

 

 

 

12.  Castor with brake DGS M F ANT 

029000011 

13.  Castor DGS M ANT 

 

029000010 

14.  Hanger 4BA-294 

 

 

604011294 

15.  Evaporative tray 

4BA-300 for DK50D 

 

604011300 

4BA-359 for DK50DM 

 

604011359 

16.  Hose flexible DN8x300 

 

072000039 

17.  Air pump 

2BA-313 for DK50D 230V/50Hz 

602011313 

2BA-314 for DK50D 230V/60Hz 

602011314 

2BA-315 for DK50D 115V/60Hz 

602011315 

4BA-330 for DK50DM 230V/50Hz  602011330 
4BA-331 for DK50DM 230V/60Hz  602011331 
4BA-332 for DK50DM 115V/60Hz  602011332 

18.  Internal back cover 4BA-301 

604011301 

19.  Solenoid valve complete  

4BA-298 230V/50-60Hz 

 

604011298 

4BA-321 115V/60Hz 

 

604011321 

20.  Powering connector GSW   

033500008 

21.  Fan NMB 4715MS  230V/50-60Hz  035300006 

Fan LFT 1212 AD 115V/60Hz 

035300005 

22.  Fast-on coupling 21 KSKS08MPC  025500129 
23.  Hose PA 8/6 

 

 

 

24.  Cooler complete 4BA-296   

604011296 

25.  Siren BZ-06 

 

 

037900007 

26.  Terminal board 307-5 

 

033100024 

27.  Filtering circuit 

1ELF 6VY for DK50D 230V  

031600001 

1ELF10VY for DK50D 115V, 
DK50DM 230V 

 

 

031600002 

1ELF 16VY for DK50DM 115V 

031600003 

28.  Termo switch TH63-80°C   

033510003 

29.  Termo switch plate 4KA-178 

062000020 

30.  Termo switch holder 4KB-553 

023000311 

31.  Termo switch TH63-40°C   

033510001 

32.  Top termo switch plate  4KA-122 

062000019 

33.  Inlet filter 4KB-552   

 

025000011 

34.  Capacitor 

not valid for DK50D 
30M/450V for DK50DM 230V/50-60Hz  031330003 
2x60M/450V for DK50DM 115V/60Hz  031330010 

35.  Gasket Cu 4KA-078 

 

025900003 

36.  Safety valve 4BA-025 

 

604011025 

37.  Gasket Cu 4KA-104 

 

025900004 

38.  Connecting nut 4KA-103 

 

024000027 

39.  Pressure switch compl. 4BA-358 

604011358 

40.  Manometer complete 4BA-271 

604011271 

41.  Humidity indicator compl. 4BA-272  604011272 
42.  Operation hours counter  

HK30.G0 230V/50Hz 

 

037200008 

HK30.G0 230V/60Hz 

 

037200010 

HK30.G0 115V/60Hz 

 

037200011 

43.  Glow tube LS-3P1 red 

 

033600001 

44.  Glow tube LS-3P1 green 

 

033600002 

45.  Tube 4BA-346 

 

 

604011346 

46.  Motor for DK50D 

1LF9073-4AJ99-ZN52 230V/50Hz  035110017 
1LF9073-4AJ99-ZN50 230V/60Hz  035110010 
1LF9073-4AJ99-ZN51 115V/60Hz   035110013 
Motor for DK50DM 
1LF7096-4AJ97-ZN50 230V/50-60Hz  035110008 
1LF7096-4AJ77-Z 115V/60Hz 

035110015

 

47.  Crank case 1KA-794 

 

050000012 

48.  Ledge 4KB-547 

 

 

023000321 

49.  Screw M8x20 

 

 

041000053 

50.  Crank 

4DA-054 for DK50D 

 

604031054 

4CA-014 for DK50DM / 50Hz 

604021014 

4CA-158 for DK50DM / 60Hz 

604021158 

51.  Bearing 6204 2RS   

 

021000001 

52.  Crank case cap 

4KB-231 for DK50D 

 

024000386 

2KA-793 for DK50DM 

 

050000013 

53.  Input filter (B) 4KA-292 

 

025000005 

54.  Prefilter (A) 4KB-565 

 

025000012 

55.  Suction plug 4KB-136 

 

062000058 

56.  Screw M4x12 

 

 

041000020 

57.  Guiding piston ring 4KA-014 

069000005 

58.  Flexibla ring 4KA-013 

 

022000001 

59.  Sealing piston ring 4KA-012 

069000004 

60.  Cylinder 

4KB-204 for DK50D 

 

050000021 

2KA-791 for DK50DM 

 

050000014 

61.  O-ring 50x2 

 

 

073000109 

62.  Washer 6   

 

 

043000007 

63.  Screw M6x110 

 

 

041000045 

64.  Cylinder head complete 4CA-070  604021070 
65.  Spring 4KA-019 

 

 

022000010 

66.  Membrane seat 3KA-015 

 

024000006 

67.  Ring 4KB-575 for DK50D / 60Hz 

023000330 

68.  Screw 4KB-123 (M6x30) 

 

041000086 

69.  Piston with piston rod 4CA-004 

604021004 

70.  Handle 4BA-242 

 

 

604011242 

71.  Upper platte  

4CA-128 for DK50D 

 

604021128 

4CA-138 for DK50DM 

 

604021138 

72.  Washer 6,6/22 

 

 

043000033 

73.  Washer 6   

 

 

043000007 

74.  Screw M6x12 

 

 

041000138 

75.  Manometer bracket 4KB-185 

023000178 

76.  Screw M4x20 

 

 

041000129 

 
 

PDF created with pdfFactory trial version 

www.pdffactory.com

Содержание DK50 D

Страница 1: ...PDF created with pdfFactory trial version www pdffactory com...

Страница 2: ...6 4 Cooling failure alarm 10 6 5 Cleaning the compressor 10 7 SERVICE 10 7 1 Intervals of service 10 7 2 Service kits 11 7 3 Preparations for service 12 7 4 Check of safety valve 12 7 5 Replacement of...

Страница 3: ...l at the time of printing corresponds to the design of the appliance and its state according to the relevant safety and technical standards The manufacturer reserves all copyrights for the given wirin...

Страница 4: ...d expert 1 5 Alert notices and symbols For your information the symbols below are used in the installation operation and maintenance manual on packaging materials and on the product Information or ins...

Страница 5: ...r 6 Main input power switch primary fuses 7 Equipotential pin 8 Socket for the supply lead 9 WALL input air coming from another pressure air source for example central distribution 10 OUT output air f...

Страница 6: ...ity Lit 5 5 Working pressure of the pump bar 5 0 7 0 5 0 7 0 Operation pressure of safety valve bar 8 0 8 0 Adjustment of the air pressure at the output pressure regulator pressure regulator Dimension...

Страница 7: ...peration for the first time by a qualified expert His obligation is to also train the operating personnel about the usage and common maintenance of the appliance Installation and training of personnel...

Страница 8: ...t OUT port In such a connection the compressor serves as back up source of the compressed air In the case of the reduction of the pressure of air from central distribution system under 5 0 bars compre...

Страница 9: ...of supporting equipment must be in the axis of the tube max 1100 mm above the floor Other connecting consult with manufacturer Producer recommends an supporting equipment disassemble before transport...

Страница 10: ...clean the product using a solution containing abrasive components chemical solvents or another aggressive chemical 7 SERVICE Repairs that exceed the framework of common maintenance must be performed...

Страница 11: ...0V 50Hz type 4EA 034 Part No 604041034 DK50 D 230V 60Hz type 4EA 030 Part No 604041030 DK50 DM 230V 60Hz type 4EA 035 Part No 604041035 DK50 D 115V 60Hz type 4EA 031 Part No 604041031 DK50 DM 115V 60H...

Страница 12: ...st be now be closed The safety valve must not be used for depressurizing the pressure tank It could endanger the function of the safety valve 7 5 Replacement of filters of air pump Caution hot surface...

Страница 13: ...the control button of the regulator pushing down Setting the pressure in the pressure regulator must only be done by a qualified expert 7 8 Replacement of pressure switch Prior the intervention to th...

Страница 14: ...the pressure regulator Check the condition of pump itself impurities in crank box freedom at crank shaft In the case of need replace faulty parts 7 11 Checking of capacity Before checking of capacity...

Страница 15: ...hten screw on connecting rod Disassemble the preparation for ensuring piston embed cylinder onto crank box seat of membrane 5 with O ring 6 2 pcs spring 4 complete cylinder head 3 and screw screws 4 p...

Страница 16: ...idity indicator is out of green zone Large consumption of compressed air Do not exceed max flow see technical data Leakage of air from pneumatic distribution system Control of pneumatic distribution s...

Страница 17: ...A 4KB 565 025000012 2 pcs Input filter B 4KA 292 025000005 6 pcs Fuse 2 pcs for DK50 D 230V T6 3A 038100004 for DK50 D 115V T10A 038100005 for DK50 DM 230V T10A 038100005 for DK50 DM 115V T16A 038100...

Страница 18: ...erminal board 8 Thermoswitch 9 Power indicator 10 Cooling failure indicator 11 Counter 12 Acoustic indicator 13 Capacitor 2 5 U2 6 PE N U 1 4 4 3 t 40 C 8 1 2 3 4 230V 50Hz 230V 60Hz 115V 60Hz TYPE B...

Страница 19: ...er 3 Compressor 4 Solenoid valve 5 Sound absorber 6 Cooler 7 Filter with the water trap 8 Check valve 9 Air tank 10 Pressure regulator with filter 11 Air Outlet 12 Air Inlet 13 Main input power switch...

Страница 20: ...SERVICE MANUAL 20 SM eVENT GB 3_8 2006 doc PDF created with pdfFactory trial version www pdffactory com...

Страница 21: ...SERVICE MANUAL SM eVENT GB 3_8 2006 doc 21 PDF created with pdfFactory trial version www pdffactory com...

Страница 22: ...SERVICE MANUAL 22 SM eVENT GB 3_8 2006 doc PDF created with pdfFactory trial version www pdffactory com...

Страница 23: ...SERVICE MANUAL SM eVENT GB 3_8 2006 doc 23 PDF created with pdfFactory trial version www pdffactory com...

Страница 24: ...031330010 35 Gasket Cu 4KA 078 025900003 36 Safety valve 4BA 025 604011025 37 Gasket Cu 4KA 104 025900004 38 Connecting nut 4KA 103 024000027 39 Pressure switch compl 4BA 358 604011358 40 Manometer co...

Страница 25: ...166 4KB 595 for DK50DM 061000177 85 Case 4BA 293 for DK50D 604011293 4BA 329 for DK50DM 604011329 86 Foam 4KB 538 061000170 87 Socket for tube 4BA 347 604011347 88 Foam 4KB 539 061000171 89 Foam 4KB 5...

Страница 26: ...SERVICE MANUAL 26 SM eVENT GB 3_8 2006 doc PDF created with pdfFactory trial version www pdffactory com...

Страница 27: ...ENT GB 3_8 2006 doc 27 EKOM spol s r o Priemyseln 5031 18 921 01 PI E ANY Slovak republic tel 421 33 7967255 fax 421 33 7967223 e mail ekom ekom sk www ekom sk PDF created with pdfFactory trial versio...

Страница 28: ...ging w x d x h mm 500x530x850 500x600x850 15 LIST OF SPARE PARTS 12 Castor with brake MKR JM S 75 029000058 13 Castor MKR S 75 029000059 CHANGE 2 Description Termo switches TH63 are changed by switche...

Страница 29: ...z 604011405 4BA 406 115V 60Hz 604011406 Previous and recent spare parts are interchangeable CHANGE 4 Description 1 Filters 111211 40B and RS 1701 became part of basic outfit 2 Change of evaporative tr...

Отзывы: