2.  INFORMATION ON PRODUCT 

2.1. 

Usage on purpose  

• 

This medical compressor, hereinafter referred to as the compressor, is the source of clean, oil-

free  pressurized  air  for  use  with  respiratory  equipment.  Due  to  the  silent  operation  of  the 
compressor, it is possible to place the unit near the respiratory appliance beside the bed of the 
patient. 

The  compressed  air  from  the  compressor  is  suitable  for  the  operation  of  respiratory  
appliances or similar apparatus.  

• 

The  compressor  is  to  be  operated  in  dry,  aerated  rooms  where  the  ambient  temperature  is 

b15°C and +40°C, with the relative air humidity not exceeding 70%.  

• 

The compressor must not be exposed to rain. The appliance must not be used in humid or wet 

environments.  Furthermore,  it  is  forbidden  to  use  it  in  the  presence  of  flammable  liquids  or 
gasses. 

• 

Prior  to  the  connection  of  compressor  to  the  respiratory appliance, it is necessary to assess 

whether the available medium meets the requirements of the given appliance of use. Please, 
refer to the technical data for this purpose. 

• 

Usage,  other  than  the  compressors  intended  use  is  not  considered  to  be  safe.  The 

manufacturer  is  not  responsible  for  the  damages  implied  if  the  compressor  is  used  for  other 
uses. Such a risk is assumed only by operator/user. 

2.2. 

Description of product  

(see annex)

(1)  PRESSURE - the output air pressure 

manometer 

(2)  HUMIDITY - the humidity indicator 

(performance) 

(3)  Cooling failure indicator 
(4)  Power indicator 
(5)  Hour meter 
(6)  Main input, power switch, primary fuses 
(7)  Equipotential pin 
(8)  Socket for the supply lead 
(9)  WALL - input air coming from another 

pressure air source (for example central 
distribution) 

(10)  OUT - output air from compressor  
(11)  Pressure tank 
(12)  Air intake filter 
(13)  Pressure regulator with filter 
(14)  Filter and water trap 
(15)  Cooler 
(16)  Pressure switch 
(17)  Safety valve 
(18)  Internal back cover 
(19)  Air pump 
(20)  Evaporative tray 
(21)  Fixative elements 
(22)  Tube for supporting equipment

 
The appliance consists of an oil-free piston compressor driven by a single-phase electric motor, of 
an  filtering  and  drying  equipment,  of  an  pressure  tank  and  of  a  pressure  regulator.  The 
compressor  is  placed  in  a  box  lined  with  noise  reducing  material.  The  appliance  thus  produces 
filtered compressed air without any traces of oil with reduced humidity contents. 
The air pump (19) sucks the atmospheric air through filters (A), (B). Compressed air is cooled in 
the  cooler  (15),  continue  in  the  filter  (C)  and  condensed  liquid  from  the  water  trap  (14)  is 
automatically  released  to  the  evaporative  tray (20). The air continue through the check valve to 
the  pressure  tank  (11)  and  through  the  pressure  regulator  (13)  with  filter (D) to the output (10). 
The fans operate while the air pump (19) works, or also during the pause, when the temperature 
in the cabinet is over 40°C. Any increase of temperature in the cabinet over 80°C (cooling failure) 
is  signalled  by  the  red  indicator  (3)  and  by  acoustic  alarm.  The  pressure  switch  (16)  control 
running  the  air  pump  (19)  according  to  pressure  in  the  pressure  tank  (11).  A  safety  valve  (17) 
prevents the maximal allowable pressure in the pressure tank. 

PDF created with pdfFactory trial version 

www.pdffactory.com

Содержание DK50 D

Страница 1: ...PDF created with pdfFactory trial version www pdffactory com...

Страница 2: ...6 4 Cooling failure alarm 10 6 5 Cleaning the compressor 10 7 SERVICE 10 7 1 Intervals of service 10 7 2 Service kits 11 7 3 Preparations for service 12 7 4 Check of safety valve 12 7 5 Replacement of...

Страница 3: ...l at the time of printing corresponds to the design of the appliance and its state according to the relevant safety and technical standards The manufacturer reserves all copyrights for the given wirin...

Страница 4: ...d expert 1 5 Alert notices and symbols For your information the symbols below are used in the installation operation and maintenance manual on packaging materials and on the product Information or ins...

Страница 5: ...r 6 Main input power switch primary fuses 7 Equipotential pin 8 Socket for the supply lead 9 WALL input air coming from another pressure air source for example central distribution 10 OUT output air f...

Страница 6: ...ity Lit 5 5 Working pressure of the pump bar 5 0 7 0 5 0 7 0 Operation pressure of safety valve bar 8 0 8 0 Adjustment of the air pressure at the output pressure regulator pressure regulator Dimension...

Страница 7: ...peration for the first time by a qualified expert His obligation is to also train the operating personnel about the usage and common maintenance of the appliance Installation and training of personnel...

Страница 8: ...t OUT port In such a connection the compressor serves as back up source of the compressed air In the case of the reduction of the pressure of air from central distribution system under 5 0 bars compre...

Страница 9: ...of supporting equipment must be in the axis of the tube max 1100 mm above the floor Other connecting consult with manufacturer Producer recommends an supporting equipment disassemble before transport...

Страница 10: ...clean the product using a solution containing abrasive components chemical solvents or another aggressive chemical 7 SERVICE Repairs that exceed the framework of common maintenance must be performed...

Страница 11: ...0V 50Hz type 4EA 034 Part No 604041034 DK50 D 230V 60Hz type 4EA 030 Part No 604041030 DK50 DM 230V 60Hz type 4EA 035 Part No 604041035 DK50 D 115V 60Hz type 4EA 031 Part No 604041031 DK50 DM 115V 60H...

Страница 12: ...st be now be closed The safety valve must not be used for depressurizing the pressure tank It could endanger the function of the safety valve 7 5 Replacement of filters of air pump Caution hot surface...

Страница 13: ...the control button of the regulator pushing down Setting the pressure in the pressure regulator must only be done by a qualified expert 7 8 Replacement of pressure switch Prior the intervention to th...

Страница 14: ...the pressure regulator Check the condition of pump itself impurities in crank box freedom at crank shaft In the case of need replace faulty parts 7 11 Checking of capacity Before checking of capacity...

Страница 15: ...hten screw on connecting rod Disassemble the preparation for ensuring piston embed cylinder onto crank box seat of membrane 5 with O ring 6 2 pcs spring 4 complete cylinder head 3 and screw screws 4 p...

Страница 16: ...idity indicator is out of green zone Large consumption of compressed air Do not exceed max flow see technical data Leakage of air from pneumatic distribution system Control of pneumatic distribution s...

Страница 17: ...A 4KB 565 025000012 2 pcs Input filter B 4KA 292 025000005 6 pcs Fuse 2 pcs for DK50 D 230V T6 3A 038100004 for DK50 D 115V T10A 038100005 for DK50 DM 230V T10A 038100005 for DK50 DM 115V T16A 038100...

Страница 18: ...erminal board 8 Thermoswitch 9 Power indicator 10 Cooling failure indicator 11 Counter 12 Acoustic indicator 13 Capacitor 2 5 U2 6 PE N U 1 4 4 3 t 40 C 8 1 2 3 4 230V 50Hz 230V 60Hz 115V 60Hz TYPE B...

Страница 19: ...er 3 Compressor 4 Solenoid valve 5 Sound absorber 6 Cooler 7 Filter with the water trap 8 Check valve 9 Air tank 10 Pressure regulator with filter 11 Air Outlet 12 Air Inlet 13 Main input power switch...

Страница 20: ...SERVICE MANUAL 20 SM eVENT GB 3_8 2006 doc PDF created with pdfFactory trial version www pdffactory com...

Страница 21: ...SERVICE MANUAL SM eVENT GB 3_8 2006 doc 21 PDF created with pdfFactory trial version www pdffactory com...

Страница 22: ...SERVICE MANUAL 22 SM eVENT GB 3_8 2006 doc PDF created with pdfFactory trial version www pdffactory com...

Страница 23: ...SERVICE MANUAL SM eVENT GB 3_8 2006 doc 23 PDF created with pdfFactory trial version www pdffactory com...

Страница 24: ...031330010 35 Gasket Cu 4KA 078 025900003 36 Safety valve 4BA 025 604011025 37 Gasket Cu 4KA 104 025900004 38 Connecting nut 4KA 103 024000027 39 Pressure switch compl 4BA 358 604011358 40 Manometer co...

Страница 25: ...166 4KB 595 for DK50DM 061000177 85 Case 4BA 293 for DK50D 604011293 4BA 329 for DK50DM 604011329 86 Foam 4KB 538 061000170 87 Socket for tube 4BA 347 604011347 88 Foam 4KB 539 061000171 89 Foam 4KB 5...

Страница 26: ...SERVICE MANUAL 26 SM eVENT GB 3_8 2006 doc PDF created with pdfFactory trial version www pdffactory com...

Страница 27: ...ENT GB 3_8 2006 doc 27 EKOM spol s r o Priemyseln 5031 18 921 01 PI E ANY Slovak republic tel 421 33 7967255 fax 421 33 7967223 e mail ekom ekom sk www ekom sk PDF created with pdfFactory trial versio...

Страница 28: ...ging w x d x h mm 500x530x850 500x600x850 15 LIST OF SPARE PARTS 12 Castor with brake MKR JM S 75 029000058 13 Castor MKR S 75 029000059 CHANGE 2 Description Termo switches TH63 are changed by switche...

Страница 29: ...z 604011405 4BA 406 115V 60Hz 604011406 Previous and recent spare parts are interchangeable CHANGE 4 Description 1 Filters 111211 40B and RS 1701 became part of basic outfit 2 Change of evaporative tr...

Отзывы: