EKO INSTRUMENTS CO., LTD. VIP Checker HC-121 Instruction Manual V2
Pg. 34
9. Specification
9-1. Specifications
Table 9-1. Specification
Characteristics
Details
Test Sample Size
Requirements
Width:
50mm difference against reference VIP (when the reference VIP is over
200x200mm)
Thickness:
5mm difference against reference VIP (when the reference VIP is over 10mm)
Measurement Conditions
Room temperature:
Within 25+/-2°C, temperature controlled
Humidity:
Within 40%+/-5%, stabilized
Air from air conditioner does not blow VIP samples directly.
No vibration, dust, high voltage equipment, high electro-magnetic and
electrostatic
Table 9-2. Each Unit Specifications
Characteristics
Details
HC-120 Measurement Unit
Sensor Input
Differential Thermocouple (Thermopile type), Output in mV
Input Range
10mV fixed
Measurement Accuracy
±0.025mV
Heater Current
160mA (per each Sensor Unit channel)
Input Channels
5 channels
Measurement Time
60sec/1ch for standard:
Standby Time:
[Standard 20sec.] (Can be changed above 20 sec.)
Cooling Time:
[Standard 30sec.] (Can be changed above30 sec.)
Heating Time:
[Standard 10sec.] (Cannot be changed)
Buzzer
Beeps at completion of a measurement.
Communication
RS-232C
PC Connection Port
RS-232C
Barcode Reader Connection
Port
BAR CODE READER
Size
320(W) x 120(H) x 220(D) mm
Weight
3kg
Power Source
AC100 - 240V, 50/60Hz, Fuse: 3A