EKO INSTRUMENTS CO., LTD. VIP Checker HC-121 Instruction Manual V2
Pg. 19
5.
After approximately 3 minutes, buzzer sounds and the start button starts to blink as the
measurement is completed.
The measured values are displayed in the specific section for the sensors and the 5x3 samples.
6.
Return the Sensor Unit on the Heatsink and push the start button. Start button stops blinking and
.
7.
When the cooling time is over, the start button lights up. Wait for more than 5 minutes, then
repeat the steps 2~6.
If the sample itself gets warmed up by the heater on the Sensor Unit, the calibration cannot be
performed well, thus wait for more than 5 minutes till the Sensor Unit cools down.
Figure 6-19. Start Measurement Window
Figure 6-20. Finish Measurement Window
30 seconds,
30 seconds, then
30 seconds, till the Sensor Unit cools down.