36
WWW.EILERSEN.COM
Load cell protocol type
401x
4x40-StdLC
2010
4x40-StdBB
2010
4x40-StdBB
2010
4x40-StdBB
2010
4x40-StdBB
Number of load cells
1-4
1
2
3
4
Sampling time
2
5
10
15
20
Filter
Filter frequency (Hz) and total settling time (ms)
Taps
Damping
7
-60dB
120 Hz
14 ms
48 Hz
35 ms
24 Hz
70 ms
16 Hz
105 ms
12 Hz
140 ms
9
-60dB
100 Hz
18 ms
40 Hz
45 ms
20 Hz
90 ms
13 Hz
135 ms
10 Hz
180 ms
9
-80dB
120 Hz
18ms
48 Hz
45ms
24 Hz
90ms
16 Hz
135ms
12 Hz
180ms
12
-60dB
80 Hz
24 ms
32 Hz
60 ms
16 Hz
120 ms
11 Hz
180 ms
8 Hz
240 ms
12
-80dB
100 Hz
24 ms
40 Hz
60 ms
20 Hz
120 ms
13 Hz
180 ms
10 Hz
240 ms
15
-80dB
80 Hz
30 ms
32 Hz
75 ms
16 Hz
150 ms
11 Hz
225 ms
8 Hz
300 ms
17
-60dB
60 Hz
34 ms
24 Hz
85 ms
12 Hz
170 ms
8 Hz
255 ms
6 Hz
340 ms
21
-80dB
60 Hz
42 ms
24 Hz
105 ms
12 Hz
210 ms
8 Hz
315 ms
6 Hz
420 ms
25
-60dB
40 Hz
50 ms
16 Hz
125 ms
8 Hz
250 ms
5 Hz
375 ms
4 Hz
500 ms
32
-80dB
40 Hz
64 ms
16 Hz
160 ms
8 Hz
320 ms
5 Hz
480 ms
4 Hz
640 ms
50
-60dB
20 Hz
100 ms
8 Hz
250 ms
4 Hz
500 ms
2,7 Hz
750 ms
2 Hz
1000 ms
64
-80dB
20 Hz
128 ms
8 Hz
320 ms
4 Hz
640 ms
2,7 Hz
960 ms
2 Hz
1280 ms
67
-60dB
15 Hz
134 ms
6 Hz
335 ms
3 Hz
670 ms
2 Hz
1005 ms
1,5 Hz
1340 ms
85
-80dB
15 Hz
170 ms
6 Hz
425 ms
3 Hz
850 ms
2 Hz
1275 ms
1,5 Hz
1700 ms
100
-60dB
10 Hz
200 ms
4 Hz
500 ms
2 Hz
1000 ms
1,3 Hz
1500 ms
1 Hz
2000 ms