Change dosing parameters
From the menu system the
DOSING PARAMETERS
screen is selected:
In this screen dosing parameters such as registration period and afterflow correction are
shown and can be changed.
Change registration period
The registration period is the time from a dosing has finished to the actual amount dosed is
registered automatically. The registration period is entered in milliseconds (0 - 30000 ms).
The value should be set large enough to ensure no vibrations and a stable weight indication is
achieved before the automatic registration is performed. A value of 0 ms indicates that
NO
automatic registration is performed, and that a registration must be made using the
REGIS-
TRATION
input (or the keyboard).
Change afterflow status
The afterflow status indicates whether afterflow correction is disabled (OFF), enabled using a
fixed correction (ON) or enabled using an automatically adjusted correction (AUTO).
Change afterflow correction
The actual afterflow correction used during do
sing if enabled using the “Afterflow status” pa-
rameter (ON or AUTO), is the amount prior to the net weight reaching the FINE limit, where
the dosing is assumed to be completed and the FINE and COARSE outputs are deactivated.
The “Afterflow” parameter ind
icates the actual correction.
Please notice:
T
he “Afterflow” parameter is not used if afterflow correction is disabled by
setting the “Afterflow status” pa
rameter to OFF.
Please notice:
T
he “Afterflow” parameter can be changed automatically after each dosin
g if
afterflow correction is performed automaticall
y by setting the “Afterflow status” parameter
to AUTO.