M2
Value measured and displayed by Querx,
second measurement
T1
Actual value (measured by reference device),
first measurement
T2
Actual value (measured by reference device),
second measurement
After calculating the required values for
Offset
and
Gain
, as described above, open the page for the
corresponding sensor in the Querx configuration area:
Sensors > Temperature / Humidity / Air Pressure
.
1.
Enter the calculated value into the input field
Offset
.
2.
Enter the calculated value into the input field
Gain
.
3.
Leave the input field
Annotation
empty.
4.
Click
Save
to apply the offset and gain to all future
measurements.
11.7.3 Automatic Adjustment with Offset and Gain
egnite Querx can calculate offset and gain
automatically, if two reference values, i.e. reliable
measurements (see above), are known. The device
applies the same formulas as detailed in the section
Manual Adjustment with Offset and Gain
.
In order to let Querx adjust offset and gain
automatically, please proceed as follows:
Open the page
Sensors > Temperature / Humidity / Air
Pressure
in the Querx configuration area.