Installation & User Manual - Sampling Systems
Version 140618 UK
Page 40
Replacing the tubing inside the pump body
-
Remove the power supply
-
Open the quick release of the peristaltic pump to access the
tubing.
-
Remove rubber band on the water detection sensor
-
Remove the tubing from the pump housing and water detection
sensor, you can turn the pump rollers to make removing the tube
easier.
-
Clean the housing of the pump with a cloth, you can remove the
pump rollers by loosening the hex crews on the rollers frame
inside the pump for good access.
-
Cut new tubing in the same length as the original tube.
-
Place the replacement tubing inside, you can turn the pump
rollers to make placing the tube easier.
-
Place the tubing inside the sensor tube.
-
Close the pump body with the quick release.
-
Restore the tubing connections.
-
Take a sample and check if everything functions.