Installation & User Manual - Sampling Systems
Version 140618 UK
Page 17
3.5.5
Changing Program Settings
Select submenu
Program Settings
and press
Enter
.
Select from the following parameters. Press
Enter
to change its value.
Parameter
Description
Start Date-time
Enter date & time to start sampling (00-00 = off).
Stop Date-time
Enter date & time to stop sampling (00-00 = off).
Stop after full containers
Enter desired number of container to be filled, the
sampler then stop sampling & restarts when reset is
pushed.
Date time.
Enter time & date. Change it when needed.
Password
Enter a number from 1-9999 to change the password.
3.5.6
Changing Input / Output Settings
Select submenu
Input / Output Settings
and press
Enter
.
Select from the following parameters. Press
Enter
to change its value.
Parameter
Description
Flow signal
Select with
◄
or
►
to totalize the desired flow signal.
Pulse = totalize contact (>50 ms) on the pulse input.
Current = totalize a 4-20mA flow signal.
Pulse + Current = totalize contact on the pulse input,
current is only to display.
Pulse input value
Select the pulse unit= 0=none, 1=m3, 2= l, 3=p
Also assign a value to the pulse contact, 0,1 -1000,0
(0,1 for current flow signal).
Current Value
Enter the 20mA value of the current signal
2600,0m
3
/h max.
Input 1 config.
Input 2 config.
Select with
◄
or
►
to configure the inputs.
Choose from the following functions:
Start/Stop PRG, Start Program, Stop program
Take Sample, Next container, Start Cool unit,
Not in Use.
Output 1 config.
Output 2 config.
Select with
◄
or
►
to config. the outputs.
Choose from the following functions:
General Alarm, Sample Alarm, Temperature High,
Sampling Active, Sample OK, Error sample, 1 m
3
pulse,
0,1 m
3
pulse, containers full.
Communication settings
ID# = Unit ID number, Asign a unique network ID
number, Pari= Parity (0=even, 1=odd, 2=none),
Baudrate (legal baudrates=300, 600, 1200,2400, 4800,
9600)
3.5.7
Changing Cool unit Settings
Select submenu
Cool Unit Settings
and press
Enter
.
Select from the following parameters. Press
Enter
to change its value.
Parameter
Description
Cool Unit On/Off
On and off set points cool unit.
Defrost cycle
Fill in the desired interval (F) in hours and how many
minutes the Cool Unit needs to defrost (t).
Temperature. offset
Temperature offset, to correct the temperature
measured inside the cabinet.
Defrost Time
Fill in the desired time (in hours) to stop the cool unit.
Set to 0 hours to activate the Cool Unit again.
ECO cool
Not available.
Содержание omy
Страница 43: ...Installation User Manual Sampling Systems Version 140618 UK Page 43 12 Menu Navigatie Jazz Controller ...
Страница 44: ...Installation User Manual Sampling Systems Version 140618 UK Page 44 ...
Страница 45: ...Installation User Manual Sampling Systems Version 140618 UK Page 45 ...
Страница 46: ...Installation User Manual Sampling Systems Version 140618 UK Page 46 ...
Страница 47: ...Installation User Manual Sampling Systems Version 140618 UK Page 47 ...
Страница 48: ...Installation User Manual Sampling Systems Version 140618 UK Page 48 ...