12
Getting Started
Edelbrock QwikData 2
Calibrating in Manual ADC Count
Follow the Manual ADC Count procedure to manually enter calibration points from the keyboard for your
sensors. This is a useful procedure if you are using a sensor supplied with specific calibration information. Be
sure to edit the selected channels parameters before creating a calibration.
1. Open the calibration window and select the channel you want to calibrate.(Fig.1)
Fig.1
2. On the calibration window menu under Channel select Manual ADC Count, a check should appear.
3. On the Calibration window Menu Bar under Calibrate, select Insert Value.
4. Enter the Y Value (Units)
Hit OK
5. Enter the X Value (Bits)
Hit OK
The graphic display is updated as points are added to the calibration grid. If calibration points exceed graph
adjust graph scaling in step #7 of
.
6. Repeat step 2 to enter more calibration points if necessary.
7. Enter a Calibration Type.
8. Close to save changes.
Содержание QwikData 2
Страница 27: ...26 Getting Started Edelbrock QwikData 2 Suggested Harness Routing...
Страница 28: ...27 Getting Started Edelbrock QwikData 2 91290 Basic Analog Harness Details...
Страница 29: ...28 Getting Started Edelbrock QwikData 2 91290 Basic Harness Schematic...
Страница 31: ...30 Getting Started Edelbrock QwikData 2 91294 Advanced Thermocouple Harness Details...
Страница 32: ...31 Getting Started Edelbrock QwikData 2 91294 Advanced Thermocouple Harness Schematic...
Страница 34: ...33 Getting Started Edelbrock QwikData 2 91291 Advanced Analog Harness Details...
Страница 35: ...34 Getting Started Edelbrock QwikData 2 91291 Advanced Analog Harness Schematic...
Страница 39: ...38 Getting Started Edelbrock QwikData 2 User Notes...