VE-H100B Veterinary Pulse Oximeter User Manual
- 61 -
synchronously flash with PR value.
5.2.6 Lower than Lo Alarm Limit
If the measured SpO
2
or PR value is lower than the
Lo Alarm
(lower alarm limit), the oximeter gives a high alarm.
Here we take SpO
2
for example:
If the measured SpO
2
value is lower than the set
SpO
2
Lo Alarm
,
the oximeter gives a low SpO
2
alarm. (Make sure
Alarm System
in menu is
ON
.)
A
↓
icon displays near SpO
2
value, which indicates the
measured value is lower than that of
SpO
2
Lo Alarm
, it will
synchronously flash with SpO
2
value.
5.2.7 Alarm Silence
If
Alarm System
in menu is
ON
, pressing the
alarm silence
button, the audio alarm will be off for the pause period set by the
user, but the visual alarm is still active.
When the audio alarm is off, press the
Alarm Silence
button to
reactivate the audio alarm function.
5.2.8 Turning off Alarm System
After the alarm system is turned off, the oximeter can not give a
visual or an audio alarm except for low battery icon alarm.
Set
Alarm System
to
ON
, the alarm system will be active. It will
give an audio alarm and a visual alarm if an alarm occurs.