VE-H100B Veterinary Pulse Oximeter User Manual
- 98 -
communications equipment.
Rated
maximum
output power
of transmitter
(W)
Separation
distance
according
to
frequency of transmitter (m)
150 kHz to
80 MHz
P
d
2
.
1
80 MHz to
800 MHz
P
d
2
.
1
800 MHz to 2.5
GHz
P
d
3
.
2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed
above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE
1
At 80 MHz and 800 MHz, the separation
distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.