higher the sensitivity is, the quicker the pulse oximeter responds to
the changes in the patient’s oxygen saturation level. Contrarily,
the lower the sensitivity is, the slower the pulse oximeter responds
to the changes in the patient’s oxygen saturation level, but the
measurement accuracy will be improved. When a critical patient is
monitored, selecting high sensitivity will help to understand the
patient’s state.
3.3.9 Alarm System
Set
Alarm System
to
ON
or
OFF
to turn on or off the alarm
system.
If
Alarm system
is set to
OFF
, a dialog box pops up as follows:
Figure 3-14 Confirm to Turn off Alarm
If
Alarm System
is
ON
and an alarm occurs, the oximeter will
give a visual alarm and an audio alarm.
Pressing the
Alarm Silence
button can suspend the alarm system
for seconds (the pause period can be set to 30, 60, 90 or 120s by
the user, see section 3.3.5), the audio alarm off icon displays. But
the visual alarm is still active. For example, if the measured SpO
2
value is higher than
SpO
2
Hi Alarm
or lower than
SpO
2
Lo