VE-H100B Veterinary Pulse Oximeter User Manual
- 74 -
4.
Leave the sensor to air dry for at least 30 minutes.
WARNING
Sterilization may cause damage to the equipment and is
therefore not recommended for this pulse oximeter unless
otherwise indicated in your hospital’s servicing schedule.
CAUTION
Never use EtO or formaldehyde for disinfection.
The replacement of accessories, such as cables, sensors etc.,
should be taken according to actual usage. It is recommended to
replace accessories once a year. Please refer to installation
methods in relevant chapters for replacement.
Production date can be found on labels. The service life for main
machine (not including replaceable accessories or parts) is 5 years
when working time is 8 hours per day. Maintenance should be
taken half a year for continuous use if service life exceeds 5 years.