SE-1200 Series Electrocardiograph User Manual Operation Preparations
- 25 -
IEC
AHA
Electrode Placement
C1
V1
Fourth intercostal space at the right border of the sternum
C2
V2
Fourth intercostal space at the left border of the sternum
C3
V3
Fifth rib between C2 and C4
C4
V4
Fifth intercostal space on the left midclavicular line
C5
V5
Left anterior axillary line at the horizontal level of C4
C6 V6
Left
midaxillary
line
at the horizontal level of C4
L
LA
Left arm/Left deltoid
R
RA
Right arm/Right deltoid
F
LL
Left leg/Upper leg as close to torso as possible
N
RL
Right leg/Upper leg as close to torso as possible
NEHB Placement
IEC
AHA
Electrode Placement
N
st
A1
Attachment point of the second rib to the right sternal edge
N
ax
A2
Fifth intercostal space on the left posterior axillary line
N
ap
V4
Left
mid-clavicular
line
in the fifth intercostal space
R RA
Right
arm
L LA
Left
arm
N or RF
RL
Right leg
F LL
Left
leg