SE-1200 Series Electrocardiograph User Manual Printing ECG Reports
- 62 -
beat from lead V5 / The maximum absolute value of the amplitude of S or S’
wave of one selected dominant beat from lead V1;
RV5+SV1
(optional)
Sum of RV5 and SV1;
RV6/SV2
(optional)
The maximum of the amplitude of R or R’ wave of one selected dominant
beat from lead V6 / The maximum absolute value of the amplitude of S or S’
wave of one selected dominant beat from lead V2;
Average Template:
Average template shows the average value of 10s sampled ECG
signals of every lead.
The broken lines on the template are position markers. They
respectively mark the start and end points of the P and QRS
waves, and the end point of the T wave.
Diagnosis Information:
Diagnosis information shows the auto diagnosis result.
Bottom Information:
0.67~100Hz (0.67Hz DFT Filter, 100Hz Lowpass Filter),
AC50 (50Hz AC Filter)
25mm/s (Paper Speed)
10mm/mV (Gain)
60 (Heart Rate)
2*5s+1r (12 leads are printed in 2 groups of 6 with the ECG wave
of one lead on the bottom, and every group is printed for about 5s)
Time Scale (0-10s)
V2.1 (Software Version)
SEMIP V1.81 (Algorithm Version; if Glasgow algorithm is
enabled, Glasgow V28.6.0 will be printed instead.)
SE-1200 (Electrocardiograph Model)
Institution Name
Report Confirmed by