SE-1200 Series Electrocardiograph User Manual Operation Preparations
- 22 -
Chapter 3 Operation Preparations
WARNING
Before use, the equipment, patient cable and electrodes should be checked. Replace
them if there is any evident defectiveness or aging which may impair the safety or the
performance, and make sure that the equipment is in proper working condition.
3.1 Connecting the Patient Cable to the Electrocardiograph
and Electrodes
WARNING
The performance and electric shock protection can be guaranteed only if the original
patient cable and electrodes of the manufacturer are used.
The patient cable includes the main cable and lead wires which can be connected to electrodes.
3.1.1 Connecting the Patient Cable to the Electrocardiograph
Connect the patient cable to the patient cable socket on the right side of the main unit, and then
secure them with two screws.
3.1.2 Connecting the Patient Cable to Electrodes
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the
reusable electrodes or the alligator clips. Firmly attach them.
The identifiers and color codes of electrode connectors used comply with IEC/EN requirements.
In order to avoid incorrect connection, the identifiers and color codes are specified in Table 3-1.
Moreover the equivalent codes according to AHA requirements are given in Table 3-1 too.
Connecting to
Electrodes
Lead Wires
Connecting to the
Patient Cable Socket
Main Cable