C3A, C6A, C6A HDVideoColposcope User Manual EMC Information
- 101 -
Appendix 4 EMC Information
A4.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
The
Video Colposcope
is intended for use in the electromagnetic environment specified
below. The customer or the user of the
Video Colposcope
should assure that it is used in
such an environment.
Emission test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The
Video Colposcope
uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emission
CISPR 11
Class A
The
Video Colposcope
is suitable for use in all
establishments, other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies