Acclarix AX3 Series Diagnostic Ultrasound System User Manual Measurements and Reports
- 83 -
No.
Cardiac Measurement
Description
Tool
Thickness at End-diastole
IVSTs
Interventricular Septal Thickness at
End-systole
LVIDs
Left Ventricular Internal Diameter at
End-systole
LVPWs
Left
Ventricular
Posterior
Wall
Thickness at End-systole
3.2
LVET
Left Ventricular Ejection Time
3.3
MV(Mitral
Valve)
E-F Slope
Mitral Valve E-F slope
EPSS
E point septal separation
3.4
LA/Ao
LA
Left atrial diameter
AoD
Aortic root diameter
RVOT Diam
Right Ventricular outflow tract
3.5
HR**
Heart rate
Table 8-4 Cardiac Measurements
* The tools can be switched by
Tool Options
in the touch screen after the Measurement is selected.
**System provides 1-8 cycles for HR measurement, depending on measurement setup.
Simpson
’s Method of Disks (MOD)
Simpson
‟s MOD is a method used to calculate LV diastolic and systolic volumes and ejection fraction.
Simpson‟s MOD is available in the Cardiac presets.
To use the Simpson
‟s MOD measurement:
1. Invoke the B-mode Measure function.
2. Select
LV Simpson
measurement on the touch screen and then select one measurement label
from
A4C Dias.
,
A4C Sys.
,
A2C Dia
s., and
A2C Sys.
.A single caliper will be displayed on the
image field.
3. Move the caliper and press <
Set>
key to position the start point A at one end of left ventricular
long axis.
4. Draw the trace along the endocardium of the left ventricle.
5. Press <
Set>
key to fix the end point B and complete the trace. A new caliper positions
automatically at the apex (Point D) of the trace, connecting the Point C (Midpoint of point A and B).
This distance between point C and D is the longest detected by the system, as shown below:
6. Pivot the Point D caliper to a new desired position if necessary.
7. Press <
Set>
key to complete the measurement.