Acclarix AX3 Series Diagnostic Ultrasound System User Manual EMC Information
- 138 -
Appendix D EMC Information
Electromagnetic emissions
Guidance and manufacture’s declaration – electromagnetic emission
The system is intended for use in the electromagnetic environment specified below; The customer or
the user of the system should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The system is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC61000-3-3
Complies
NOTE:
The EMISSIONS characteristics of the system make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is
normally required) the system might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or re-orienting the
equipment.
Содержание Acclarix AX15
Страница 1: ...1 ...
Страница 127: ...Acclarix AX3 Series Diagnostic Ultrasound System User Manual Utilities 121 Figure 11 10 Screen Adjust Screen ...
Страница 149: ... 143 ...