Acclarix AX3 Series Diagnostic Ultrasound System User Manual Transducers and Biopsy
- 57 -
3. Properly secure the clamp of bracket with the locking knob. Ensure the backet is firmly attached,
and then place the biopsy needle into the needle guide path.
BGK-001/BGK-003
The installation steps for BGK-001 and BGK-003 brackets are the same. Here we take one bracket for
illustration.
Structures:
Installation and Use Steps:
1. Place an appropriate amount of gel on transducer surface, and insert transducer into the sheath.
2. Loosen the locking knob to open the clamp bracket. Attach the bracket to the transducer by
aligning the locating markers on the bracket and the transducer. Properly secure the clamp of
bracket with the locking knob.
Needle Guide Holder
Clamp
Locking Knob