9
Table 1.1 (continued)
Name
Unit of
measurement
Standardized
value
Main relative permissible error limit of
measurement of the effective specific
activity of naturally occurring
radionuclides
226
Ra,
232
Th,
40
K in the
objects of the environment in
geometries “2
”, “4
”, Р=0.95
%
35
Measurement range of gamma radiation
ambient dose equivalent rate from
137
Cs,
134
Cs
μSv/h
0.001 - 10
Main relative permissible error limits
of measurement of gamma radiation
ambient dose equivalent rate from
137
Cs,
134
Cs, given Р=0.95
%
20
Unstable readings of the radiometer
during 8 hours of operation, not more
than
%
5
Measurement range of gamma radiation
ambient dose equivalent rate
)
10
(
*
H
μSv/h
0.01 – 50.00
Main relative permissible error limit of
measurement of ambient dose
equivalent rate (DER) of gamma
radiation, given P=0.95
%
15+1/
)
10
(
*
H
,
where
)
10
(
*
H
– is
a numeric value
of gamma
radiation DER,
μSv/h
Sensitivity to gamma radiation for
137
Cs,
134
Cs, not less than
(pulse/s) /(μSv/h)
500
Energy range of registered gamma
radiation
-
50 keV –
3.0 MeV
Energy dependence of readings for
energies 59 keV (
241
Аm),
1.3 MeV (
60
Со) and 0.66 MeV (
137
Cs)
%
30
Содержание VIRTUOSO RKG-14
Страница 1: ...RKG 14 VIRTUOSO multipurpose environment activity radiometer Operating manual I 412139 003 02...
Страница 2: ......
Страница 4: ...2 Version 8 Last updated June 2021...
Страница 76: ...74 a b Figure 2 59 Track recording Figure 2 60 Track pausing...
Страница 98: ...96 b VIRTUOSO measurement report...
Страница 99: ...97 VIRTUOSO measurement report...