Instruction manual CCS 5, 15/ 10/ 2021
18
•
The measurement
mode selected in the
preceding menu, the
classification type is
shown in the headline.
•
Both the water
saturation and the
temperature are shown
with each
measurement. When a
fluid type is also
selected, then the ppm
water content is also
displayed.
•
The numbers of each
measurement and the
date of start is
displayed.
•
The measurements are
continued until they get
stopped with button
[
STOP
].
•
To continue the
measurement just press
the button [START], The
numbers of each
measurement and the
date of start is
displayed.
•
The last measurement results will not be deleted by continuing the measurement.
To assure constant mode of operation of the unit and the accuracy of measurement
results it is recommendable to perform a flushing of the CCS 5 with cleaned mineral
oil (filtrated H22) for some minutes after finishing the measurements.