Instruction manual CCS 5, 15/ 10/ 2021
16
After entering
all the
necessary
measurement
parameters,
confirm with
and
this will take
you to the
next menu.
2.5.2.1.
Measure-type: single
•
This mode is to accomplish a single measurement.
•
The procedure as described in detail below is performed in any selected report standard
(ISO 4406:99, NAS 1638, SAE AS 4059).
•
Start the
measurements
with [
Start
].
The flush process runs before the measurement, it takes from 15
seconds to 300 seconds, based on the selection in the setup
menu.
•
The operating mode selected, the report standard type and
the measuring mode are shown in the headline
•
Contamination
class.
•
Both the water
saturation and the
temperature are
shown with each
measurement.
When a fluid type
is selected then
the ppm water
content is also
displayed.
•
Particle concentration per 1 ml