2.3 Load Setting Operation
Through the system menu, you can enter the load setting interface, as shown in Figure 2.3. In
this interface, the relevant settings for the load range, limit value, delay off, etc. can be
completed.
Operation instructions: 1. Select the operation item by turning the knob. 2. Non-numeric
parameters Press [Enter] to switch the setting options. 3. Digital parameter, press [Enter] to
enter the edit mode, use the arrow keys to select the corresponding number of digits, then
rotate the knob to adjust the value, [Enter] to confirm the input. 4. Press [Shift] + [
▶
] (Esc)
to return to the previous interface.
Fig - 2.3 Interface of Load Setting
2.4 Basic Operation
Programmable electronic Load can work under these basic measurement mode:
Constant current mode (CC), constant voltage mode (CV), constant resistance mode (CR),
constant power mode (CP), constant voltage mode (CC+CV), constant resistance mode
(CC+CR). All these mode could be set through button of
【
Mode
】【
Enter
】【
Esc
】
on the testing display.
Fig -2.4 Interface of Mode Selection
2.4.1 Constant Current Measurement Mode