24
7 Advanced Application
7.1 Setting 0 % and 100 % Output
Parameters
As for stepping operation and percentage display, 0% and 100% should be set before
using. The default values have been set when delivered from the factory and set values
are listed below:
Source function
0% value
100% value
DC V
0.000 V
10.000 V
DC mV
0.00 mV
100.00 mV
DC mA
4.000 mA
20.000 mA
Resistance 400Ω
0.00 Ω
400.00 Ω
Resistance 3200Ω
0.0 Ω
3200.0 Ω
Frequency 200Hz
0.00 Hz
200.00 Hz
Frequency 2,000Hz
0.0 Hz
2000.0 Hz
Frequency 20 kHz
0.000 kHz
20.000 kHz
TC - J model
0.0
℃
1000.0
℃
TC - K model
0.0
℃
1000.0
℃
TC - T model
0.0
℃
400.0
℃
TC - E model
0.0
℃
800.0
℃
TC - R model
0
℃
1500
℃
TC - S model
0
℃
1500
℃
TC - B model
600
℃
1800
℃
TC - N model
0.0
℃
1000.0
℃
Pt100
0.0
℃
500.0
℃
Pt1000
0.0
℃
400.0
℃
Cu50
0.0
℃
150.0
℃
Cu100
0.0
℃
150.0
℃
The default set values may not meet your requirements, so you can reset them. Press
and hold
or
to reset 0% and 100% value until the buzzer works. The
reset value will be stored in the storage space of the calibrator automatically and
remains effective after restart. Now you can start operation with the reset value:
Manually stepping an output with 25 % increments.
Jump between the 0 and 100 % span points by momentarily pushing
or
.