20
Figure 5.1.2-1
2.
If you need to adjust the manual compensation value, press
keys
to adjust.
3.
Press
key to save the value of manual compensation and exit from the
setup mode at the same time, reading as shown in Figure 5.1.2-2.
Figure 5.1.2-2
4.
If necessary, press
key return to automatic compensation.
5.2 Using Resistance Thermometer Detector
(RTD)
The calibrator accepts RTD types of Pt100, Pt1000, Cu50 and Cu100.The calibrator
accepts RTD measurement inputs in two-, three-, or four-wire connections, with the
three-wire connection the most common. A four-wire configuration provides the
highest measurement precision, and two-wire provides the lowest measurement
precision.
To measure temperature using an RTD input, proceed as follows: