Er317
Instrument non-volatile memory un-initialized. Return to Dynisco.
Er900
ROM error. Return to Dynisco.
Er901
RAM error. Return to Dynisco.
Er902
Key stuck detected. Check domed buttons for damage.
Er903
CPU fault. Return to Dynisco.
ooooo
Over-range indication.
This status is displayed when the A/D converter value is out of range, or the input signal is
greater than full scale value plus 27% of span, or the displayed value exceeds the display
capability of 99900.
-oooo
Under-range indication.
This status is displayed when the A/D converter value is out of range, or the input signal is less
than full scale value minus 27% of span, or the displayed value exceeds the display capability of
-1990.
OPEn
This message is displayed for an open voltage or thermo input, current reads 0.