Chapter 3 System Design
28
Agility System
Operator’s Manual
3.4 Sample Pipettor Arm
3.4.1 Overview
The sample pipettor (Figure 3-4) arm performs the following operations:
Gets a new sample tip from the sample tip rack
Ejects used sample tips into tip waste container evenly
Transfers sample fluid from sample tube to the microplate wells and/or dilution strips
Transfers dilution fluid from the dilution strips to microplate wells
Transfers controls or standard fluids from reagent packs to microplate wells or dilution
strips
Mixes fluid within a dilution strip
Figure 3-4: Sample Pipettor
The sample pipettor is capable of dispensing between 10 µL and 300 µL in single shot mode
with an accuracy of < 2% of target volume in its operating range (single shot) based on using
a new tip and a pipetting precision of < 3% CV at 10 shots at any volume in operating range
based on single shot with a new tip (both specifications based on using de-ionized water)
Содержание Agility Automated ELISA System
Страница 6: ...Table of Contents 6 Agility System Operator s Manual This page intentionally left blank ...
Страница 8: ...About this Manual 8 Agility System Operator s Manual This page intentionally left blank ...
Страница 20: ...Chapter 1 Overview 20 Agility System Operator s Manual Figure 1 17 USB Hub Setup Figure 1 18 USB Hub Setup ...
Страница 24: ...Chapter 2 Safety 24 Agility System Operator s Manual This page intentionally left blank ...
Страница 42: ...Chapter 3 System Design 42 Agility System Operator s Manual This page intentionally left blank ...
Страница 52: ...Chapter 4 Operating the Agility System 52 Agility System Operator s Manual Figure 4 9 Wash Assay Screen ...
Страница 74: ...Chapter 4 Operating the Agility System 74 Agility System Operator s Manual Figure 4 28 Consumables Status ...
Страница 82: ...Chapter 4 Operating the Agility System 82 Agility System Operator s Manual This page intentionally left blank ...
Страница 98: ...Chapter 6 The Graphical User Interface 98 Agility System Operator s Manual Figure 6 1 The Administrator Screen ...
Страница 101: ...Chapter 6 The Graphical User Interface Agility System Operator s Manual 101 Figure 6 4 The Version Report Screen ...
Страница 144: ...Chapter 7Error Handling 144 Agility System Operator s Manual This page intentionally left blank ...
Страница 148: ...Appendix A Specifications 148 Agility System Operator s Manual This page intentionally left blank ...
Страница 150: ...Appendix B Consumables and Spare Parts 150 Agility System Operator s Manual This page intentionally left blank ...